| Error message |
Describes the situation |
Solution |
| E-b/ Err 36 |
The batteries are insufficiently power to operate the meter. The meter will not operate until replace or charge the batteries. |
You have to replace or charge the batteries soon. |
| E-C/ Err 22 |
Invalid code strip insertion. E.g. get wrong check/code strip, defective check/code strip. |
Only use the same brand of code strip which were produced specifically for your glucose monitor. |
| E-U/ Err 24 |
Inserts a used test strip. |
Repeat the test with a new test strip. |
“ ” Blood sample is input before the symbol for blood blinks. |
Wait for the “ ” symbol to blink then re-test using proper amount of blood. |
| E-t/ Err25 |
The meter or test strip appear when environmental temperature is below system operate range. |
System operation range is 10~40℃ (50~104℉). Repeat the test after the meter or test strip have reached the above temperature. |
| E-t/ Err 26 |
The meter or test strip appear when environmental temperature is above system operate range. |
| E-F |
Remove the strip after applying blood to the absorbent hole. |
Repeat the test with a new test strip and wait for test result. |
| Lo |
Poor amount of blood sample or blood glucose value under measurement range 20 mg/dL (1.1mmol/l). |
1.Sufficient amount of blood sample and repeat the test with a new test strip.
2.Blood glucose measurement range is 20~600 mg/dL (1.1~33.3 mmol/L).
3. Repeat the test and seek immediate medical attention. |
| Hi |
Blood glucose value over measurement range 600 mg/dL (33.3mmol/L). |
| Err.01 |
Cuff serious leakage or inflation too slow. |
1.Check cuff is fitted correctly, and also refit cuff tightly and correctly to measure blood pressure again and don’t move your arm during the measurement.
2. If the error message appears again, you will contact with our distributor or us. |
| Err.05 |
Deflation is too fast. |
| Err.06 |
Deflation is too slow. |
| Err.07 |
Systolic outsides the legal range. |
| Err.08 |
Mean pressure outsides the legal range. |
| Err.02 |
Cannot figure out systolic. |
1.Check cuff is fitted correctly, and also refit cuff tightly and correctly to measure blood pressure again and don’t move your arm during the measurement.
2. If the error message appears again, you will contact with our distributor or us. |
| Err.04 |
Cannot figure out diastolic or it outsides the legal range. |
| Err.09 |
Cuff pressure is not enough even if the highest value is pumped. |
| Err.03/ Err.330 |
Auto inflation cuff didn’t stop until pulsation disappear and more elevated than 300mmHg. |
| Other error message |
The meter may some defective problems. |
Repeat the test with a new test strip or repeat measurement. If the error message appears again, you will contact with our distributor or us. |